www.restek.it

Cannabis Potency Compounds on Rxi-35Sil MS
Peaks | |
---|---|
1. | Cannabichromene![]() |
2. | Cannabidiol![]() |
3. | delta-8-Tetrahydrocannabinol![]() |
Peaks | |
---|---|
4. | delta-9-Tetrahydrocannabinol![]() |
5. | Cannabigerol![]() |
6. | Cannabinol![]() |

Column | Rxi-35Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13820) |
Sample | Cannabinoids standard (cat.# 34014) |
Cannabichromene (cat.# 34092) | |
delta-8-Tetrahydrocannabinol (THC) (cat.# 34090) | |
Cannabigerol (cat.# 34091) | |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 225 °C (hold 0.1 min) to 330 °C at 35 °C/min (hold 0.9 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 2.5 mL/min |
Detector | FID @ 350 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.