www.restek.it

QuEChERS Extract of Cannabis on Rxi-17Sil MS and Rxi-5ms by GCxGC-TOFMS (Zoomed in on terpenes region)

Column | Rxi-17Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123) |
Rxi-5ms 0.9 m, 0.18 mm ID, 0.18 µm, | |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 20:1) |
Liner: | Premium 4 mm Precision Inlet Liner w/Wool (cat.# 23305.1) |
Inj. Temp.: | 275 °C |
Oven | |
Oven Temp.: | Rxi-17Sil MS: 40 °C (hold 1 min) to 320 °C at 5 °C/min (hold 3 min) Rxi-5ms: 45 °C (hold 1 min) to 325 °C at 5 °C/min (hold 3 min) |
Carrier Gas | He, corrected constant flow (2 mL/min) |
Modulation | |
Modulator Temp. Offset: | 20 °C |
Second Dimension Separation Time: | 2 sec |
Hot Pulse Time: | 0.6 sec |
Cool Time between Stages: | 0.4 sec |
Detector | TOFMS |
Transfer Line Temp.: | 310 °C |
Analyzer Type: | TOF |
Source Temp.: | 250 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 4.5 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45-500 amu |
Spectral Acquisition Rate: | 200 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | Sample Preparation: Sample: Up to 2 g of finely ground cannabis in a 50 mL polypropylene centrifuge tube. Extraction: Add 10 mL acetonitrile and 10 mL of organic-free water, then shake thoroughly. Soak 1 hour, then aggressively vortex 30 minutes. Add Q-sep packet (cat.# 26235), shake 1 minute, then centrifuge 5 minutes at 3,000 g with Q-sep 3000 centrifuge (cat.# 26230). |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.