Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Comparison of Strawberry Extracts (TICs) Processed with Different QuEChERS dSPE Sorbents on Rxi-5Sil MS

image
GC_FF1199
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleStrawberries processed with QuEChERS sample preparation
QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267)
anthracene for quality control (cat.# 33264)
Diluent:acetonitrile
Injection
Inj. Vol.:1 µL splitless (hold 1.5 min)
Liner:Gooseneck Splitless (5mm) (cat.# 22973-200.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1.5 min) to 340 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorMS
Mode:
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesSample Preparation:
Sample: 10 g of homogenized strawberries, fortified with pesticides and QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267).

Extraction: 10 mL acetonitrile added, 1 minute shake, then addition of Q-sep Q110 (cat.# 26235), 1 minute shake, centrifuge with Q-sep 3000 Centrifuge (cat.# 26230).

dSPE: Cleanup procedure according to EN 15662 method, add control standard anthracene (cat.# 33264) to 1 mL extract, add this to one of two different Q-sep dSPE tubes (either cat.# 26215 or cat.# 26217 ), shake, centrifuge.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

pesticide residues

,

dispersive SPE

,

SPE

,

fruits

,

produce

,

food safety

,

dispersive solid phase extract

,

PSA

,

GCB

,

cleanup

,

extraction

,

EN 15662

,

Q-sep

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.