Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Carfentrazone Ethyl in Dandelion Root Powder (Dietary Supplement) on Rxi-5Sil MS

image
GC_FF1172
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sampledandelion root powder spiked with 46 pesticides
QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267)
anthracene for quality control (cat.# 33264)
Diluent:toluene
Conc.:80 pg/µL
Injection
Inj. Vol.:1 µL splitless (hold 1.5 min)
Liner:5mm Splitless with wool (cat.# 22975-200.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1.5 min) to 340 °C at 8 °C/min
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorMS
Mode:
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec.
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesPurge valve time: 1.5 min.
Detector mode: full mass range

Sample Preparation:
Wetting: 1 g of dandelion root powder combined with 9 mL of water, shaken well, fortified with pesticides and QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267), soak for 2 hours.

Extraction: 10 mL acetonitrile added then addition of Q-sep Q110 (cat.# 26213), centrifuge with Q-sep 3000 Centrifuge (cat.# 26230).

cSPE: add control standard anthracene (cat.# 33264) to 1 mL extract, add magnesium sulfate to cartridge, use 6 mL Combo SPE Cartridge containing 500 mg CarboPrep 90/500 mg PSA (cat.# 26194), concentrate via evaporation.

relativo alle ricerche nella libreria dei cromatogrammi

botanical

,

natural products

,

nutraceuticals

,

pesticide residues

,

GMP

,

herbal

,

QuEChERS

,

SPE

,

tof

,

time of flight

,

Food Drug and Cosmetic Act

,

DSHEA

,

EN 15562

,

wetting

,

dry

,

CarboPrep 90

,

PSA

,

anthracene

,

magnesium sulfate

,

GCB

,

LECO

,

GC-TOFMS

,

. Dispersive SPE

,

Rxi-5SilMS

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us