Pesticides in QuEChERS Extract of Sweet Potato on Rxi-5Sil MS

Column | Rxi®-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602) |
Sample | sweet potato spiked with pesticide mix, extracted with acetonitrile and Q-sep™ Q110 QuEChERS extraction tube (cat.# 26213) |
Diluent: | Acetonitrile |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Gooseneck Splitless (4mm) w/Wool (cat.# 22405) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 72.5 °C (hold 1 min) to 350 °C at 20 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | TOFMS |
Transfer Line Temp.: | 300 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 3 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Scan Range: | 45-550 amu |
Scan Rate: | 10 scans/sec |
Instrument | LECO Pegasus III GC-TOFMS |
Notes | Sample: A. extract (without cleanup step) acidified with formic acid to pH 5 B. extract with cleanup using Q-sep™ Q250 QuEChERS dSPE clean-up tube (cat.# 26124), acidified with formic acid to pH 5 m/z plotted: 60, 73, 87, 129, 256 |