www.restek.it

Chlorinated Pesticide Residues in Olive Oil Rtx-CLPesticides2

Column | Rtx-CLPesticides2, 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323) |
Sample | Olive oil spiked with Organochlorine Pesticide Mix AB # 3 (cat.# 32415) |
Conc.: | 10 µg/mL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 0.5 min) |
Liner: | Single Taper w/Wool (cat.# 22286-200.1) |
Inj. Temp.: | 225 °C |
Oven | |
Oven Temp.: | 140 °C (hold 0.5 min) to 268 °C at 20 °C/min to 290 °C at 3 °C/min to 330 °C at 20 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1 mL/min |
Detector | MS |
Mode: | SIM |
Transfer Line Temp.: | 320 °C |
Ionization Mode: | EI |
Notes | Extraction and dSPE Cleanup for Pesticide Residues in Olive Oil Test sample: A 1.5 mL sample of commercially obtained virgin olive oil was spiked with a standard organochlorine pesticide mix. The spiked sample was processed as follows: 1. Dilute with 1.5 mL hexane. 2. Add 6 mL of acetonitrile (ACN). 3. Mix for 30 minutes on a shaker. 4. Allow layers to separate (approximately 20 minutes), then collect the top (ACN) layer. 5. Repeat the liquid-liquid extraction (steps 2–4) and combine both ACN extract layers. 6. Place 1 mL of the combined ACN extract in a 1.5 mL tube containing 150 mg magnesium sulfate and 50 mg PSA. 7. Shake the tube for 2 minutes. 8. Centrifuge at 3,000 U/min for approximately 5 minutes. 9. Remove the top layer and inject directly into the gas chromatograph system. |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.