5% Fragrance Materials Association Mix on Rtx-1701

PeaksConc.
(wt.%)
1.Ethyl butyrate35.7
2.Limonene20.0
3.Eucalyptol0.5
4.Geraniol0.6
5.Benzoic acid1.0
6.Cinnamic aldehyde0.5
7.Hydroxycitronellal5.0
8.Thymol0.3
9.Cinnamyl alcohol0.3
10.Cinnamyl acetate0.3
11.Vanillin0.1
12.Benzyl salicylate35.7
image
GC_FF00658
ColumnRtx-1701, 60 m, 0.25 mm ID, 0.25 µm (cat.# 12026)
SampleFragrance materials test mix (cat.# 31807)
Conc.: 5% FMA mix in acetone
Injection
Inj. Vol.:1 µL split (split ratio 40:1)
Liner:Splitless (4 mm ID) (cat.# 20814)
Inj. Temp.:285 °C
Oven
Oven Temp.:50 °C to 270 °C at 3 °C/min
Carrier GasHe, constant flow
Flow Rate:0.6 mL/min
DetectorFID @ 300 °C