Fatty Acids (Free) on Stabilwax-DA (30m, 0.32mm ID, 0.25µm)

Peaks
1.C4:0 Butyric acid
2.C6:0 Caproic acid
3.C8:0 Caprylic acid
4.C10:0 Capric acid
5.C12:0 Lauric acid
6.C14:0 Myristic acid
7.C16:0 Palmitic acid
8.C18:0 Stearic acid
9.C18:1 Oleic acid
10.C18:2 Linoleic acid
11.C18:3 Linolenic acid
12.C20:0 Arachidic acid
13.C22:0 Behenic acid
image
GC_FF00653
ColumnStabilwax-DA, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11024)
SampleFree fatty acid mix
Diluent:Methanol
Conc.:100 ppm
Injection
Inj. Vol.:1 µL splitless (hold 0.25 min)
Liner:Laminar cup splitter
Inj. Temp.:250 °C
Purge Flow:75 mL/min
Oven
Oven Temp.:40 °C to 250 °C at 10 °C/min (hold 15 min)
Carrier GasH2, constant flow
Flow Rate:6 mL/min
DetectorFID @ 250 °C