Cleaning Solvents on Rtx-Wax

Peaks
1.Methanol
2.Isopropanol
3.Ethanol
4.n-Propanol
5.Isobutanol
6.n-Butanol
7.Methyl Cellosolve®
Peaks
8.Limonene
9.Cellosolve®
10.Butyl Cellosolve®
11.Dipropylene glycol monomethyl ether (DPGMME)
12.Diethylene glycol ethyl ether (DEGEE)
13.Ethylene glycol
14.Tripropylene glycol monomethyl ether (TPGMME)
15.Phenyl Cellosolve®
image
GC_FF00644
ColumnRtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424)
Sample1% each solvent in methanol
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Splitless (4mm ID) (cat.# 20814)
Inj. Temp.:250 °C
Oven
Oven Temp.:35 °C (hold 4 min) to 85 °C at 3 °C/min to 200 °C at 15 °C/min (hold 2 min)
Carrier GasH2, constant flow
Flow Rate:3 mL/min
DetectorFID @ 250 °C
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

solvents

,

cleaning solvents

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us