Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Flavor Compounds in Grappa on Stabilwax-DA

Peaks
1.Acetic acid
2.Ethyl decanoate
3.Ethyl dodecanoate
4.2-Phenylethanol
5.Octanoic Acid
6.Decanoic acid
7.Dodecanoic acid
8.Dibutyl phthalate
image
GC_FF00558
ColumnStabilwax-DA, 30 m, 0.18 mm ID, 0.18 µm (cat.# 550752)
Injection
Inj. Vol.:10 µL splitless
Oven
Oven Temp.:60 °C (hold 2 min) to 100 °C at 20 °C/min to 240 °C at 5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:45 cm/sec
DetectorMS
Mode:Scan
Transfer Line Temp.:240 °C
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Ionization Mode:EI
Scan Range:30-400 amu
NotesStd. conc.: neat
Gerstel CIS Injector: 35 °C (hold 2 min), to 300 °C @ 10 °C/sec. (hold 5 min)
Helium vent flow: 600 mL/min with 1.8 min vent end time
AcknowledgementChromatogram courtesy of Kevin MacNamara, Ph.D., Irish Distilleries, Ltd.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.