Fast Dual Column Analysis of Organochlorine Pesticides (EPA Method 8081) Using a GC Accelerator Kit and Rtx-CLPesticides and Rtx-CLPesticides2 Column Set
Peaks | Conc. (ng/mL) | |
---|---|---|
1. | Tetrachloro-m-xylene![]() | 25 |
2. | α-BHC![]() | 25 |
3. | γ-BHC![]() | 25 |
4. | β-BHC![]() | 25 |
5. | δ-BHC![]() | 25 |
6. | Heptachlor![]() | 25 |
7. | Aldrin![]() | 25 |
8. | Heptachlor epoxide![]() | 25 |
9. | trans-Chlordane![]() | 25 |
10. | cis-Chlordane![]() | 25 |
11. | 4,4'-DDE![]() | 50 |

GC_EV1498
Columns | Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141-051) |
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324-051) | |
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039) | |
with universal angled “Y” Press-Tight connector (cat.# 20403-261) | |
Sample | Organochlorine pesticide mix AB #2 (cat.# 32292) |
Pesticide surrogate mix (cat.# 32453) | |
Diluent: | n-Hexane |
Injection | |
Inj. Vol.: | 2 µL splitless (hold 0.3 min) |
Liner: | Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 120 °C (hold 0.3 min) to 200 °C at 60 °C/min to 230 °C at 25 °C/min to 320 °C at 35 °C/min (hold 1.5 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 90 cm/sec |
Detector | Micro-ECD @ 330 °C |
Make-up Gas Flow Rate: | 60 mL/min |
Make-up Gas Type: | N2 |
Data Rate: | 50 Hz |
Instrument | Agilent 7890A GC |
Notes | A GC Accelerator oven insert kit (cat.# 23849) was installed in the instrument to speed up analysis. The analytical columns were custom products that were not on cages. |