Fast Dual Column Analysis of Organochlorine Pesticides (EPA Method 8081) Using a GC Accelerator Kit and Rtx-CLPesticides and Rtx-CLPesticides2 Column Set

PeaksConc.
(ng/mL)
1.Tetrachloro-m-xylene25
2.α-BHC25
3.γ-BHC25
4.β-BHC25
5.δ-BHC25
6.Heptachlor25
7.Aldrin25
8.Heptachlor epoxide25
9.trans-Chlordane25
10.cis-Chlordane25
11.4,4'-DDE50
PeaksConc.
(ng/mL)
12.Endosulfan I50
13.Dieldrin50
14.Endrin50
15.4,4'-DDD50
16.Endosulfan II50
17.4,4'-DDT50
18.Endrin aldehyde50
19.Methoxychlor250
20.Endosulfan sulfate50
21.Endrin ketone50
22.Decachlorobiphenyl50
image
GC_EV1496
ColumnsRtx-CLPesticides 30 m, 0.25 mm ID, 0.25 µm (cat.# 11123-051)
and Rtx-CLPesticides2 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323-051)
using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029)
with universal angled “Y” Press-Tight connector (cat.# 20403-261)
SampleOrganochlorine pesticide mix AB #2 (cat.# 32292)
Pesticide surrogate mix (cat.# 32453)
Diluent:n-Hexane
Injection
Inj. Vol.:2 µL splitless (hold 0.3 min)
Liner:Topaz 4 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:120 °C (hold 0.3 min) to 200 °C at 60 °C/min to 230 °C at 25 °C/min to 320 °C at 35 °C/min (hold 1.5 min)
Carrier GasH2, constant flow
Linear Velocity:100 cm/sec
DetectorMicro-ECD @ 330 °C
Make-up Gas Flow Rate:60 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesA GC Accelerator oven insert kit (cat.# 23849) was installed in the instrument to speed up analysis. The analytical columns were custom products that were not on cages.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Accelerator,

,

Organochlorine

,

Pesticides,

,

8081

,

CLPesticides,

,

CLPesticides2

,

fast

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.