EPA Method 8061A Phthalates on Rtx-440

Column | Rtx-440, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12923) | ||||||||
Sample | EPA Method 8061A phthalate esters mixture (cat.# 33227) | ||||||||
Benzyl benzoate (cat.# 31847) | |||||||||
Diluent: | Methylene chloride | ||||||||
Conc.: | 80 µg/mL EPA 8061A phthalates (20 μg/mL internal standard) | ||||||||
Injection | |||||||||
Inj. Vol.: | 1 µL split (split ratio 20:1) | ||||||||
Liner: | Premium 3.5 mm Precision liner w/wool (cat.# 23320.1) | ||||||||
Inj. Temp.: | 280 °C | ||||||||
Oven | |||||||||
Oven Temp.: | 150 °C (hold 1 min) to 300 °C at 20 °C/min (hold 5 min) | ||||||||
Carrier Gas | He, constant linear velocity | ||||||||
Linear Velocity: | 50 cm/sec, 22.9 psi, 157.9 kPa @ 150 °C | ||||||||
Detector | MS | ||||||||
Mode: | Scan | ||||||||
Scan Program: |
| ||||||||
Transfer Line Temp.: | 300 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 280 °C | ||||||||
Electron Energy: | 70 eV | ||||||||
Solvent Delay Time: | 2 min | ||||||||
Tune Type: | PFTBA | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Shimadzu 2010 GC & QP2010+ MS | ||||||||
Notes | The linear velocity is 50 cm/sec, which is approximately 1.75 mL/min at 150 °C. The MS scan interval is 0.1 sec. |