Select Polycyclic Aromatic Hydrocarbons (PAHs) on Rxi-PAH (30 m x 0.25 mm ID x 0.10 µm df )

PeaksConc.
(µg/mL)
1.Naphthalene1.0
2.2-Methylnaphthalene0.80
3.1-Methylnaphthalene0.80
4.2,6-Dimethylnaphthalene0.80
5.Acenaphthylene1.0
6.Acenaphthene-d1020
7.Acenaphthene1.0
8.2,3,5-Trimethylnaphthalene0.80
9.Fluorene1.0
10.Dibenzothiophene0.80
11.Phenanthrene-d1020
12.Phenanthrene1.0
13.Anthracene1.0
14.1-Methylphenanthrene0.80
15.Fluoranthene1.0
16.Pyrene1.0
17.Benzidine1.0
18.3,3'-Dimethylbenzidine1.0
19.Benz[a]anthracene1.0
20.Chrysene-d1220
PeaksConc.
(µg/mL)
21.Cyclopenta[cd]pyrene0.20
22.Triphenylene1.0
23.Chrysene1.0
24.3,3'-Dichlorobenzidine1.0
25.5-Methylchrysene0.20
26.Benzo[b]fluoranthene1.0
27.Benzo[k]fluoranthene1.0
28.Benzo[j]fluoranthene1.0
29.Benzo[a]pyrene-d121.0
30.Benzo[a]pyrene1.0
31.3-Methylcholanthrene0.80
32.Dibenz(a,h)acridine0.80
33.Dibenz[a,j]acridine0.80
34.Indeno[1,2,3-cd]pyrene1.0
35.Dibenz[a,h]anthracene1.0
36.Benzo[ghi]perylene1.0
37.7H-Dibenzo[c,g]carbazole0.80
38.Dibenzo[a,l]pyrene0.20
39.Dibenzo[a,e]pyrene1.0
40.Dibenzo[a,i]pyrene1.0
41.Dibenzo[a,h]pyrene1.0
0.010 to 1.0 ng on column
image
GC_EV1397
ColumnRxi-PAH, 30 m, 0.25 mm ID, 0.10 µm (cat.# 49318)
SampleMethod 525.2 internal standard mix (cat.# 31825)
Triphenylene (CS S-13711U1)
Benzo[a]pyrene-d12 (CIL DLM-258-0)
Custom PAH SIM spike mix (cat.# 557484)
PAH native stock solution (Wellington PAH-STK-A)
PAH supplement mix for Method 8100 (cat.# 31857)
Diluent:Dichloromethane
Injection
Inj. Vol.:0.5 µL split (split ratio 10:1)
Liner:2 mm Precision liner w/deactivated wool (cat.# 20823)
Inj. Temp.:275 °C
Oven
Oven Temp.:110 °C (hold 0.57 min) to 175 °C at 65 °C/min to 210 °C at 45 °C/min to 295 °C at 5.3 °C/min to 350 °C at 35 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
11.05102, 128 25
22.08115, 141, 142, 156 25
32.5875.9, 80, 82.3, 152, 153, 154, 155, 162, 165, 170 15
43.4989, 94, 139, 152, 160, 178, 183, 188 15
54.4194.5, 165, 192 25
65.35100.9, 202 25
76.8792, 184 25
88.29106, 212 25
99.63113, 114, 120.1, 226, 228, 240.1 25
1010.57154, 251.9 25
1111.40119.7, 242 25
1213.16124.9, 126, 252 25
1314.96112.8, 118, 125.9, 132, 252, 264 25
1416.09126.1, 252, 268 25
1517.82138, 138.9, 139.2, 276, 278, 279 25
1619.57137.9, 267, 276 25
1720.77150, 302 25
1821.57149.9, 151, 302 25
Transfer Line Temp.:330 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9 mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Solvent Delay Time:1.05 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
NotesPerforms the separation of two critical sets of isobars:
1. Cyclopenta[c,d]pyrene, triphenylene, and chrysene
2. Benzo[b]fluoranthene, benzo[k]fluoranthene, benzo[j]fluoranthene
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

benzo[j]fluoranthene

,

chrysene

,

triphenylene

,

isobaric

,

isobar

,

benzofluoranthenes

,

sediment

,

environmental forensics

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us