Innovative Chromatography Products
www.restek.it
www.restek.it

Florida Beach Tarball 1 on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)
Peaks | tR(sec) | |
---|---|---|
1. | C20![]() | 468.8 |
2. | C30![]() | 694.6 |
3. | C40![]() | 856.8 |

GC_EV1387
Column | Rxi-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905) |
Sample | Tarball 1 - Florida Gulf beach |
Diluent: | Carbon disulfide |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 17.5 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.75 mL/min |
Detector | FID @ 420 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | Actual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc). |
Acknowledgement | Susan Forsyth - Walton County, FL - Tarball sample |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.