DRO/ORO Calibration Standard on Rxi-1HT (15 m x 0.25 mm x 0.10 µm)
Peaks | tR (sec) | |
---|---|---|
1. | C10![]() | 118.7 |
2. | C12![]() | 195.2 |
3. | C14![]() | 269.6 |
4. | C16![]() | 337.6 |
5. | C18![]() | 399.2 |

GC_EV1366
Column | Rxi-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950) |
Sample | DRO/ORO calibration standard (cat.# 31831) |
Diluent: | Methylene chloride |
Conc.: | 25,000 ng/µL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 87.5 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.1 min) to 400 °C at 20 °C/min (hold 1.9 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.75 mL/min |
Detector | FID @ 420 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | Actual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc). |