Polycyclic Aromatic Hydrocarbons (PAHs) in a Spring Creek Sediment Sample by GCxGC-TOFMS

GC_EV1356
Column | Rxi-PAH 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317) |
Rxi-1HT 1 m, 0.25 mm ID, 0.10 µm (cat.# 13950) | |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1.5 min) |
Liner: | Premium 4 mm single taper w/wool (cat.# 23303.5) |
Inj. Temp.: | 275 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | Rxi-PAH: 80 °C (hold 1.5 min) to 320 °C at 3 °C/min (hold 18.5 min) Rxi-1HT: 85 °C (hold 3.17 min) to 320 °C at 3 °C/min (hold 18.5 min) |
Carrier Gas | He, corrected constant flow (1.4 mL/min) |
Modulation | |
Modulator Temp. Offset: | 20 °C |
Second Dimension Separation Time: | 1.5 sec |
Hot Pulse Time: | 0.45 sec |
Cool Time between Stages: | 0.35 sec |
Detector | TOFMS |
Transfer Line Temp.: | 320 °C |
Analyzer Type: | TOF |
Source Temp.: | 300 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 7 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Spectral Acquisition Rate: | 200 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | The Rxi-1HT column (cat.# 13950) is a 15 m column. A 1 m segment was cut off and used here as the second dimension column. |
Acknowledgement | Sample extract provided by Frank Dorman, Melinda Pham; The Pennsylvania State University |