Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

25 µg/L 8260 MegaMix Revised on Rtx-VMS

Peaks
1.Diethyl ether
2.1,1-Dichloroethene
3.Trichlorotrifluoroethane
4.Carbon disulfide
5.Iodomethane
6.Allyl chloride
7.Methylene chloride
8.trans-1,2-Dichloroethene
9.Acetonitrile
10.Chloroprene
11.1,1-Dichloroethane
12.Acrylonitrile
13.cis-1,2-Dichloroethene
14.2,2-Dichloropropane
15.Bromochloromethane
16.Chloroform
17.Carbon tetrachloride
18.Methyl acrylate
19.Tetrahydrofuran
20.Dibromofluoromethane (SS)
21.1,1,1-Trichloroethane
22.1,1-Dichloropropene
23.Benzene
24.Propionitrile
25.Methacrylonitrile
26.Pentafluorobenzene (IS)
27.1,2-Dichloroethane
28.Isobutyl alcohol*
29.Trichloroethene
30.1,4-Difluorobenzene (IS)
31.Dibromomethane
32.1,2-Dichloropropane
33.Bromodichloromethane
34.Methyl methacrylate
35.1,4-Dioxane*
36.2-Chloropropanol*
37.cis-1,3-Dichloropropene
38.Toluene-d8 (SS)
39.Toluene
40.2-Nitropropane
41.Tetrachloroethene
Peaks
42.trans-1,3-Dichloropropene
43.1,1,2-Trichloroethane
44.Ethyl methacrylate
45.Dibromochloromethane
46.1,3-Dichloropropane
47.1,2-Dibromoethane
48.Chlorobenzene-d5 (IS)
49.Chlorobenzene
50.Ethylbenzene
51.1,1,1,2-Tetrachloroethane
52.m-Xylene
53.p-Xylene
54.o-Xylene
55.Styrene
56.Bromoform
57.Isopropylbenzene
58.1-Bromo-4-fluorobenzene (BFB) (IS)
59.cis-1,4-Dichloro-2-butene
60.n-Propylbenzene
61.Bromobenzene
62.1,1,2,2-Tetrachloroethane
63.2-Chlorotoluene
64.1,3,5-Trimethylbenzene
65.1,2,3-Trichloropropane
66.trans-1,4-Dichloro-2-butene
67.4-Chlorotoluene
68.tert-Butylbenzene
69.1,2,4-Trimethylbenzene
70.sec-Butylbenzene
71.p-Isopropyltoluene
72.1,3-Dichlorobenzene
73.1,4-Dichlorobenzene-d4 (IS)
74.1,4-Dichlorobenzene
75.n-Butylbenzene
76.1,2-Dichlorobenzene
77.1,2-Dibromo-3-Chloropropane
78.Nitrobenzene
79.Hexachlorobutadiene
80.1,2,4-Trichlorobenzene
81.Naphthalene
82.1,2,3-Trichlorobenzene
*Note: these compounds were not detected at 25 ppb when the sample was analyzed via unheated purge.
image
GC_EV1352
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample8260B MegaMix calibration mix, revised (cat.# 30152)
8260 Internal standard mix (cat.# 30074)
8260 Surrogate mix (cat.# 30073)
Diluent:Reverse-osmosis water
Conc.:25 µg/L (5 mL sample)
Injectionpurge and trap split
Liner:Premium 1 mm Split (cat.# 23333.1)
Purge and Trap
Instrument:EST Encon
Trap Type:Vocarb 3000
Purge: 11 min, flow 40 mL/min
Dry Purge: 3 min, flow 40 mL/min
Desorb Preheat Temp.:255 °C
Desorb: 1.0 min @ 260 °C, flow 30.9 mL/min
Bake: 6 min @ 260 °C
Interface Connection:injection port
Transfer Line Temp.:150 °C
Oven
Oven Temp.:45 °C (hold 4.5 min) to 100 °C at 12 °C/min to 240 °C at 25 °C/min (hold 1.32 min)
Carrier GasHe, constant flow
Flow Rate:0.9 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.747-3005.4
22.935-3005.2
Transfer Line Temp.:230 °C
Analyzer Type:Quadrupole
Source Type:Inert
Drawout Plate:6mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:1.5 min
InstrumentAgilent 7890A GC & 5975C MSD
NotesSome purge-and-trap instruments show increased activity of dehydrohalogenate pentachloroethane during analysis, forming tetrachloroethene (also known as perchloroethylene, PCE). Symptoms of this problem are poor matching of PCE levels between sources or a PCE single analyte test. This revised 8260 MegaMix standard is identical to the original, minus pentachloroethane, and was made for testing that does not required reporting of pentachloroethane.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.