Innovative Chromatography Products
www.restek.it
www.restek.it

PAHs in Sediment Extract on Rxi-PAH (Sampled From Spring Creek, PA)
Peaks | |
---|---|
1. | Benzo[b]fluoranthene![]() |
2. | Benzo[k]fluoranthene![]() |
3. | Benzo[j]fluoranthene![]() |
4. | Benzo[a]fluoranthene![]() |
5. | Benzo[e]pyrene![]() |
6. | Benzo[a]pyrene![]() |
7. | Perylene![]() |

GC_EV1331
Column | Rxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317) |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 300 °C |
Split Vent Flow Rate: | 20 mL/min |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 350 °C at 3.7 °C/min (hold 21 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2 mL/min |
Detector | TOFMS |
Transfer Line Temp.: | 320 °C |
Analyzer Type: | TOF |
Source Temp.: | 300 °C |
Electron Energy: | 70 eV |
Mass Defect: | 0 mu/100 u |
Solvent Delay Time: | 6 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Spectral Acquisition Rate: | 5 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Acknowledgement | Frank Dorman and Melinda Pham of The Pennsylvania State University for providing the sediment extract. |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.