1,4-Dioxane in Drinking Water (5 pg On-Column) on Rxi-624Sil MS (Extracted Ion Chromatogram)
Peaks
1.
Tetrahydrofuran-d8 (IS)
2.
Co-extracted material
Peaks
3.
1,4-Dioxane-d8 (SS)
4.
1,4-Dioxane
5.
Co-extracted material
GC_EV1263
Column
Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029)
with Universal Press-Tight connectors (cat.# 20429)
Sample
Tetrahydrofuran-d8 (cat.# 30112)
1,4-Dioxane-d8 (cat.# 30614)
1,4-Dioxane (cat.# 30287)
Diluent:
Dichloromethane
Conc.:
0.5 pg/µL (IS/SS 10 pg/µL) in extract
Injection
Inj. Vol.:
10 µL splitless (hold 1 min)
Liner:
Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:
120 °C
Purge Flow:
80 mL/min
Oven
Oven Temp.:
35 °C (hold 1 min) to 120 °C at 12 °C/min (hold 1 min)
Carrier Gas
He, constant flow
Flow Rate:
1.4 mL/min
Linear Velocity:
30.556 cm/sec @ 35 °C
Detector
MS
Mode:
SIM
SIM Program:
Group
Start Time (min)
Ion(s) (m/z)
Dwell (ms)
1
5.0
46,78,80
50
2
5.85
96,88,64,62,58
40
Transfer Line Temp.:
280 °C
Analyzer Type:
Quadrupole
Source Temp.:
230 °C
Quad Temp.:
150 °C
Solvent Delay Time:
5.0 min
Tune Type:
BFB
Ionization Mode:
EI
Instrument
Agilent 7890A GC & 5975C MSD
Notes
A drinking water sample fortified with 1,4-dioxane and surrogate standard was extracted using a Resprep activated coconut charcoal SPE cartridge (cat.# 26032) following EPA Method 522 protocol. Immediately after solvent elution, the extract was spiked with internal standard and brought up to final volume. The extract was then dried with anhydrous magnesium sulfate (this was a deviation from the method, which calls for sodium sulfate).
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.
I cookies consentono varie funzionalità che renderanno più agevole la tua visita al sito Restek. Utilizzando questo sito, acconsenti all’uso dei cookies in accordo con le nostre linee guida. Per saperne di più, vedi la nostra politica sulla privacy.
Restek Domestic Customer Service
Grazie.
Il messaggio è stato inviato. La ricontatteremo non appena possibile.
Messaggio non inviato.
Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037