Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

TCDFs on Rtx-Dioxin2 Column

PeakstR (min)
1.2,3,4,7-TCDF18.58
2.2,3,4,8-TCDF18.58
3.2,3,7,8-TCDF18.91
PeakstR (min)
4.3,4,6,7-TCDF19.10
5.2,3,6,7-TCDF19.24
6.1,2,6,9-TCDF19.24
7.1,2,3,9-TCDF19.66
image
GC_EV1252
ColumnRtx-Dioxin2, 60 m, 0.25 mm ID, 0.25 µm (cat.# 10758)
SampleLaboratory prepared test mix
Injection
Inj. Vol.:1.0 µL splitless
Inj. Temp.:290 °C
Oven
Oven Temp.:180 °C (hold 1 min) to 235 °C at 45 °C/min (hold 1 min) to 250 °C at 3 °C/min (hold 15 min) to 300 °C at 50 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.7 mL/min
DetectorMicromass Autospec Ultima
Transfer Line Temp.:290 °C
Analyzer Type: Magnetic Sector
Source Temp.:290 °C
Tune Type:PFK
Ionization Mode:EI
InstrumentAgilent/HP6890 GC
AcknowledgementCourtesy of Maxxam Analytics (Ontario, Canada).
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

chlorinated dibenzofurans

,

PCDF

,

TCDF

,

Furan

,

2330

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us