Semivolatile Organics on Rxi-5Sil MS by U.S. EPA Method 8270
Peaks
1. 1,4-Dioxane-d8 (IS)
2. 1,4-Dioxane
3. N-Nitrosodimethylamine
4. Pyridine
5. 2-Fluorophenol (SS)
6. Phenol-d6 (SS)
7. Phenol
8. Aniline
9. Bis(2-chloroethyl) ether
10. 2-Chlorophenol
11. 1,3-Dichlorobenzene
12. 1,4-Dichlorobenzene-D4 (IS)
13. 1,4-Dichlorobenzene
14. Benzyl alcohol
15. 1,2-Dichlorobenzene
16. 2-Methylphenol
17. Bis(2-Chloroisopropyl)ether
18. 3-Methylphenol
19. 4-Methylphenol
20. N-Nitrosodi-N-propylamine
21. Hexachloroethane
22. Nitrobenzene-D5 (SS)
23. Nitrobenzene
24. Isophorone
25. 2-Nitrophenol
26. 2,4-Dimethylphenol
27. Bis(2-chloroethoxy)methane
28. Benzoic acid
29. 2,4-Dichlorophenol
30. 1,2,4-Trichlorobenzene
31. Naphthalene-D8 (IS)
32. Naphthalene
33. 4-Chloroaniline
34. Hexachlorobutadiene
35. 4-Chloro-3-methylphenol
36. 2-Methylnaphthalene
37. 1-Methylnaphthalene
38. Hexachlorocyclopentadiene
39. 2,4,6-Trichlorophenol
40. 2,4,5-Trichlorophenol
41. 2-Fluorobiphenyl (SS)
42. 2-Chloronaphthalene
43. 2-Nitroaniline
44. 1,4-Dinitrobenzene
45. Dimethyl phthalate
46. 1,3-Dinitrobenzene
47. 2,6-Dinitrotoluene
Peaks
48. 1,2-Dinitrobenzene
49. Acenaphthylene
50. 3-Nitroaniline
51. Acenaphthene-d10 (IS)
52. Acenaphthene
53. 2,4-Dinitrophenol
54. 4-Nitrophenol
55. 2,4-Dinitrotoluene
56. Dibenzofuran
57. 2,3,5,6-Tetrachlorophenol
58. 2,3,4,6-Tetrachlorophenol
59. Diethyl phthalate
60. 4-Chlorophenyl phenyl ether
61. Fluorene
62. 4-Nitroaniline
63. 4,6-Dinitro-2-methylphenol
64. N-nitrosodiphenylamine (as diphenylamine)
65. 1,2-Diphenylhydrazine (as azobenzene)
66. 2,4,6-Tribromophenol (SS)
67. 4-Bromophenyl phenyl ether
68. Hexachlorobenzene
69. Pentachlorophenol
70. Phenanthrene-D10 (IS)
71. Phenanthrene
72. Anthracene
73. Carbazole
74. di-n- Butyl phthalate
75. Fluoranthene
76. Benzidine
77. Pyrene-d10 (SS)
78. Pyrene
79. p- Terphenyl-d14 (SS)
80. 3,3'-Dimethylbenzidine
81. Butyl benzyl phthalate
82. Bis(2-ethylhexyl) adipate
83. 3,3'-Dichlorobenzidine
84. Benz[a]anthracene
85. Chrysene-D12 (IS)
86. Bis(2-ethylhexyl)phthalate
87. Chrysene
88. Di-n- octyl phthalate
89. Benzo[b]fluoranthene
90. Benzo[k]fluoranthene
91. Benzo[a]pyrene
92. Perylene-D12 (IS)
93. Indeno[1,2,3-cd]pyrene
94. Dibenz[a,h]anthracene
95. Benzo[ghi]perylene
GC_EV1245
Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) 8270 MegaMix (cat.# 31850) 8270 Benzidines mix (cat.# 31852) Benzoic acid (cat.# 31879) 1,4-Dioxane (cat.# 31853) Revised B/N surrogate mix (cat.# 31888) Acid surrogate mix (4/89 SOW) (cat.# 31063) Revised SV internal standard mix (cat.# 31886)
Dichloromethane 8 µg/mL (IS/SS 20 μg/mL) 1 µL pulsed splitless (hold 0.59 min) Premium 4 mm single taper w/wool (cat.# 23303.5) 270 °C 30 psi (206.8kPa) 0.64 min 100 mL/min 40 °C (hold 1 min) to 280 °C at 25 °C/min to 320 °C at 5 °C/min (hold 1 min) He, constant flow 1.2 mL/min 39.723 cm/sec @ 40 °C MS Scan 280 °C Quadrupole 276 °C 150 °C 2.19 min DFTPP EI 35-550 amu 5.36 scans/sec Agilent 7890A GC & 5975C MSD 7890 Siltek treated EZ Twist Top split/splitless injection port (cat.# 22178) Gold-plated dual Vespel ring inlet seal (cat.# 21241)