www.restek.it

PCBs and BFRs on Rxi-XLB and Rxi-17Sil MS (GCxGC)

Column | Rxi-XLB 30 m, 0.25 mm ID, 0.25 µm (cat.# 13723) |
Rxi-17Sil MS 1 m, 0.15 mm ID, 0.15 µm (cat.# 43820) with IP deactivated guard column 0.2 m, 0.15 mm ID | |
Sample | PBDE mix (cat.# 33098) |
PCB congener mix, Method 8082A (cat.# 32416) | |
Diluent: | Isooctane |
Conc.: | 250 pg/µL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1.0 min) |
Liner: | Premium 4 mm single taper w/wool (cat.# 23303.5) |
Inj. Temp.: | 250 °C |
Purge Flow: | 20 mL/min |
Oven | |
Oven Temp.: | Rxi-XLB: 80 °C (hold 1 min) to 120 °C at 10 °C/min to 300 °C at 3 °C/min Rxi-17Sil MS: 85 °C (hold 1 min) to 125 °C at 10 °C/min to 305 °C at 3 °C/min |
Carrier Gas | He, corrected constant flow (2.0 mL/min) |
Modulation | |
Modulator Temp. Offset: | 20 °C |
Second Dimension Separation Time: | 3.5 sec |
Hot Pulse Time: | 1.25 sec |
Cool Time between Stages: | 0.50 sec |
Detector | µ-ECD @ 325 °C |
Data Rate: | 50 Hz |
Instrument | Agilent/HP6890 GC |
Notes | The Rxi-17Sil MS column (cat.# 43820) is a 10 m column. A 1 m section was used as a second dimension column. Guard column was connected using a Siltek-treated universal Press-Tight connector (cat.# 20480). |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.