www.restek.it

Pesticides and Brominated Flame Retardants by EPA Method 527 on Rxi-5Sil MS

Column | Rxi-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620) |
using IP Deactivated Guard Column 5 m, 0.53 mm ID (cat.# 10045) | |
with Universal Press-Tight Connectors (cat.# 20429) | |
Sample | PBDE Mix (cat.# 33098) |
Pesticides Mix #1, Method 527 (cat.# 33007) | |
Pesticides Mix #2, Method 527 (cat.# 33008) | |
Internal Standard, Method 527 (cat.# 33010) | |
Surrogate Standard, Method 527 (cat.# 33009) | |
Diluent: | Ethyl acetate:methylene chloride (1:1) |
Conc.: | 40 pg/µL (500 pg on-column) |
Injection | |
Inj. Vol.: | 12.5 µL splitless (hold 0.583 min) |
Liner: | Gooseneck Splitless (4mm) w/Semivolatiles Wool (cat.# 20799-231.5) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.60 min) to 320 °C at 30 °C/min (hold 1.07 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2 mL/min |
Detector | MS |
Mode: | |
Transfer Line Temp.: | 290 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | -70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 3.33 min |
Acquisition Range: | 45 - 750 amu |
Spectral Acquisition Rate: | 10 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | Carrier Gas Flow: 2 mL/min. corrected constant flow via pressure ramps |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.