Pesticides & Herbicides U.S. EPA Method 508.1 Rtx-CLPesticides & Rtx-CLPesticides2

Columns | Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) |
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141) | |
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039) | |
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261) | |
Sample | 50 ng/mL 508.1 calibration mix #1 (cat.# 32094) |
100 ng/mL 508.1 calibration mix #2 (cat.# 32095) | |
100 ng/mL 508.1 calibration mix #3 (cat.# 32096) | |
50 ng/mL pentachloronitrobenzene (cat.# 32091) | |
250 ng/mL 4,4'-dibromobiphenyl (cat.# 32092) | |
500 ng/mL atrazine (cat.# 32208) | |
500 ng/mL simazine (cat.# 32236) | |
Diluent: | Ethyl acetate |
Injection | |
Inj. Vol.: | 2 µL splitless (hold 0.75 min) |
Liner: | Cyclo double taper (4 mm) (cat.# 20896) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 80 °C (hold 0.5 min) to 155 °C at 19 °C/min (hold 1 min) to 210 °C at 4 °C/min to 310 °C at 25 °C/min (hold 0.5 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 26 cm/sec |
Detector | Micro-ECD @ 325 °C |
Notes | This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector. |