Pesticides of U.S. EPA Method 505 on Rtx-CLPesticides and Rtx-CLPesticides2
Peaks | |
---|---|
1. | Hexachlorocyclopentadiene![]() |
2. | Hexachlorobenzene![]() |
3. | Simazine![]() |
4. | Atrazine![]() |
5. | γ-BHC![]() |
6. | Heptachlor![]() |
7. | Alachlor![]() |
8. | Aldrin![]() |
Peaks | |
---|---|
9. | Heptachlor epoxide![]() |
10. | trans-Chlordane*![]() |
11. | trans-Nonachlor![]() |
12. | cis-Chlordane*![]() |
13. | Dieldrin![]() |
14. | Endrin![]() |
15. | cis-Nonachlor![]() |
16. | Methoxychlor![]() |

Columns | Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141) |
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) | |
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039) | |
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261) | |
Sample | 200 ng/mL 505 organohalide pesticide mix (cat.# 32024) |
4.2 µg/mL simazine (cat.# 32236) | |
4.2 µg/mL atrazine (cat.# 32208) | |
Diluent: | Methanol |
Injection | |
Inj. Vol.: | 2 µL splitless (hold 0.75 min) |
Liner: | Cyclo double taper (4 mm) (cat.# 20896) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 90 °C (hold 1 min) to 310 °C at 10 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 40 cm/sec |
Detector | Micro-ECD @ 325 °C |
Instrument | Agilent/HP6890 GC |
Notes | This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector. |