Polycyclic Aromatic Hydrocarbons Rxi-5Sil MS

Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) | ||||
Sample | PAH mix, 1 µL of 0.005 µg/mL (IS 2 µg/mL) | ||||
SV calibration mix #5 / 610 PAH mix (cat.# 31011) | |||||
1-Methylnaphthalene (cat.# 31283) | |||||
2-Methylnaphthalene (cat.# 31285) | |||||
2-Fluorobiphenyl (cat.# 31091) | |||||
Conc.: | 5 pg on-column | ||||
Injection | |||||
Inj. Vol.: | 1.0 µL pulsed splitless (hold 0.15 min) | ||||
Liner: | Drilled Uniliner (hole near top) w/wool (cat.# 21055-200.5) | ||||
Inj. Temp.: | 300 °C | ||||
Pulse Pressure: | 20 psi (137.9kPa) | ||||
Pulse Time: | 0.2 min | ||||
Purge Flow: | 60 mL/min | ||||
Oven | |||||
Oven Temp.: | 50 °C (hold 0.5 min) to 290 °C at 25 °C/min to 320 °C at 5 °C/min | ||||
Carrier Gas | He, constant flow | ||||
Flow Rate: | 1.4 mL/min | ||||
Detector | MS | ||||
Mode: | SIM | ||||
SIM Program: |
| ||||
Transfer Line Temp.: | 290 °C | ||||
Ionization Mode: | EI |