www.restek.it

Polycyclic Aromatic Hydrocarbons on Rxi-17

Column | Rxi-17, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13523) |
Sample | SV Calibration Mix #5 / 610 PAH Mix (cat.# 31011) |
PAH Supplement Mix for Method 8100 (cat.# 31857) | |
1-methylnaphthalene (cat.# 31283) | |
2-methylnaphthalene (cat.# 31285) | |
Conc.: | 50 µg/mL PAH mix |
Injection | |
Inj. Vol.: | 1.0 µL pulsed splitless (hold 0.2 min) |
Liner: | Drilled Uniliner (hole near top) (cat.# 21055) |
Inj. Temp.: | 300 °C |
Pulse Pressure: | 20 psi (137.9kPa) |
Pulse Time: | 0.3 min |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 100 °C (hold 0.5 min) to 320 °C at 4 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
Mode: | Scan |
Analyzer Type: | Quadrupole |
Solvent Delay Time: | 4.0 min |
Tune Type: | DFTPP |
Ionization Mode: | EI |
Scan Range: | 50-550 amu |
Instrument | HP6890 GC & 5973 MSD |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.