Innovative Chromatography Products
www.restek.it
www.restek.it

1,4-Dioxane, 2,4-Dinitrophenol, and Pyridine US EPA Method 8270D on Rxi-1ms
Peaks | |
---|---|
1. | 1,4-Dioxane![]() |
2. | 2,4-Dinitrophenol![]() |
3. | Pyridine![]() |

GC_EV00830
Column | Rxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323) |
Sample | 8270 MegaMix (cat.# 31850) |
Benzoic acid (cat.# 31879) | |
8270 Benzidines Mix (cat.# 31852) | |
Acid Surrogate Mix (4/89 SOW) (cat.# 31025) | |
Revised B/N Surrogate Mix (cat.# 31887) | |
1,4-Dioxane (cat.# 31853) | |
Conc.: | 0.5 µg/mL each analyte (0.5ng on column) |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 10:1) |
Liner: | Drilled Uniliner (hole near bottom) (cat.# 20771) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 50 °C (hold 0.5 min) to 245 °C at 25 °C/min to 330 °C at 6 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 280 °C |
Solvent Delay Time: | 2 min |
Tune Type: | DFTPP |
Ionization Mode: | EI |
Scan Range: | 35-550 amu |
Instrument | HP6890 GC & 5973 MSD |
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.