Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Organophosphorus Pesticides US EPA Method 8140/8141/8141A on Rtx-35

Peaks
1.Dichlorvos
2.Hexamethylphosphoramide
3.Mevinphos
4.Trichlorfon
5.Tributyl phosphate (SS)
6.Demeton-o
7.TEPP
8.Thionazin
9.Ethoprop
10.Sulfotepp
11.Naled
12.Phorate
13.Dicrotophos
14.Demeton-S
15.Terbufos
16.monocrotophos
17.Diazinon
18.Fonophos
19.Disulfoton
20.Dioxathion
21.Dimethoate
22.Phosphamidon isomer
23.Dichlorofenthion
24.Chlorpyrifos methyl
25.Phosphamidon
26.Ronnel
Peaks
27.Methyl parathion
28.Aspon
29.Trichloronate
30.Chlorpyrifos
31.Fenitrothion
32.Merphos
33.Malathion
34.Parathion-ethyl
35.Fenthion
36.Chlorfenvinphos
37.Crotoxyphos
38.merphos oxone
39.Prothiofos
40.Stirofos
41.Ethion
42.Sulprofos
43.Fensulfothion
44.Carbofenothion
45.Famphur
46.Triphenyl phosphate (SS)
47.Epn
48.Phosmet
49.Leptophos
50.Tri-o-cresyl phosphate
51.Azinphos-methyl
52.Azinphos-ethyl
53.Coumaphos
image
GC_EV00696
ColumnRtx-35, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10424)
Sampletriphenylphosphate (cat.# 32281)
tributylphosphate (cat.# 32280)
8140/8141 OP Pesticide Calibration Mix A (cat.# 32277)
8141 OP Pesticide Calibration Mix B (cat.# 32278)
Conc.: 0.1µg/mL US EPA Method 8141A Custom Standard Mixes (100ng/mL)
Injection
Inj. Vol.:0.5 µL splitless (hold 5 min)
Liner:Drilled Uniliner (hole near top) (cat.# 21054-214.1)
Inj. Temp.:220 °C
Oven
Oven Temp.:100 °C to 180 °C at 10 °C/min (hold 2 min) to 300 °C at 18 °C/min (hold 10 min)
Carrier GasHe, constant pressure
Linear Velocity:42 cm/sec @ 60 °C
DetectorFPD @ 280 °C
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.