Organophosphorus Pesticides US EPA Method 8140/8141/8141A on Rtx-35

Column | Rtx-35, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10424) |
Sample | triphenylphosphate (cat.# 32281) |
tributylphosphate (cat.# 32280) | |
8140/8141 OP Pesticide Calibration Mix A (cat.# 32277) | |
8141 OP Pesticide Calibration Mix B (cat.# 32278) | |
Conc.: | 0.1µg/mL US EPA Method 8141A Custom Standard Mixes (100ng/mL) |
Injection | |
Inj. Vol.: | 0.5 µL splitless (hold 5 min) |
Liner: | Drilled Uniliner (hole near top) (cat.# 21054-214.1) |
Inj. Temp.: | 220 °C |
Oven | |
Oven Temp.: | 100 °C to 180 °C at 10 °C/min (hold 2 min) to 300 °C at 18 °C/min (hold 10 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 42 cm/sec @ 60 °C |
Detector | FPD @ 280 °C |