Organophosphorus Pesticides U.S. EPA Method 8141A on Rtx-OPPesticides2

Peaks
1.Dichlorvos
2.Hexamethylphosphoramide
3.Mevinphos
4.Trichlorfon
5.TEPP
6.Demeton-o
7.Thionazin
8.Tributyl phosphate (IS)
9.Ethoprop
10.Naled
11.Sulfotepp
12.Phorate
13.Dicrotophos
14.Demeton-S
15.Monocrotophos
16.Terbufos
17.Dimethoate
18.Dioxathion
19.Fonophos
20.Diazinon
21.Disulfoton
22.Phosphamidon isomer
23.Dichlorofenthion
24.Chlorpyrifos methyl
25.Phosphamidon
26.Methyl parathion
Peaks
27.Ronnel
28.Fenitrothion
29.Aspon
30.Malathion
31.Chlorpyrifos
32.Trichloronate
33.Parathion-ethyl
34.Fenthion
35.Merphos
36.Chlorfenvinphos
37.Crotoxyphos
38.Stirofos
39.Prothiofos
40.Merphos oxone (breakdown product)
41.Ethion
42.Fensulfothion
43.Sulprofos
44.Carbophenothion
45.Famphur
46.Triphenyl phosphate (SS)
47.Epn
48.Phosmet
49.Leptophos
50.Tri-o-cresyl phosphate
51.Azinphos-methyl
52.Azinphos-ethyl
53.Coumaphos
image
GC_EV00602
ColumnRtx-OPPesticides2, 30 m, 0.25 mm ID, 0.25 µm (cat.# 11243)
SampleTriphenylphosphate (cat.# 32281)
Tributylphosphate (cat.# 32280)
8140/8141 OP pesticide calibration mix A (cat.# 32277)
8141 OP pesticide calibration mix B (cat.# 32278)
Conc.:100 ppm (100 ng on-column)
Injection
Inj. Vol.:1.0 µL splitless (hold 0.4 min)
Liner:Double taper splitless (4 mm) (cat.# 20785)
Inj. Temp.:250 °C
Oven
Oven Temp.:80 °C (hold 0.5 min) to 140 °C at 20 °C/min to 210 °C at 4 °C/min (hold 1 min) to 280 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Ionization Mode:EI
Scan Range:35-400 amu
NotesU.S. EPA Method 8141A custom standard mix. Additional mixes not shown. Contact Restek for more information.