Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Polynuclear Aromatic Hydrocarbons US EPA Method 610 on Rtx-CLPesticides2

Peaks
1.Naphthalene
2.Acenaphthylene
3.Acenaphthene
4.Fluorene
5.Phenanthrene
6.Anthracene
7.Fluoranthene
8.Pyrene
Peaks
9.Benz[a]anthracene
10.Chrysene
11.Benzo[b]fluoranthene
12.Benzo[k]fluoranthene
13.Benzo[a]pyrene
14.Indeno[1,2,3-cd]pyrene
15.Dibenzo[a,h]anthracene
16.Benzo[ghi]perylene
image
GC_EV00596
ColumnRtx-CLPesticides2, 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323)
SampleSV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
Diluent:Methylene chloride
Conc.:50 µg/mL
Injection
Inj. Vol.:1.0 µL splitless (hold 2 min)
Liner:Auto SYS Splitless (2mm ID) w/Deact. Wool (cat.# 20829)
Inj. Temp.:330 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 250 °C at 25 °C/min to 265 °C at 5 °C/min to 300 °C at 25 °C/min (hold 4 min)
Carrier GasH2, constant flow
Flow Rate:4 mL/min
DetectorFID @ 350 °C
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

PAH

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.