Organophosphorus Pesticides US EPA Method 8140/8141/8141A on Rtx-1

Column | Rtx-1, 30 m, 0.32 mm ID, 0.50 µm (cat.# 10139) |
Sample | Triphenylphosphate (cat.# 32281) |
Tributylphosphate (cat.# 32280) | |
8140/8141 OP Pesticide Calibration Mix A (cat.# 32277) | |
8141 OP Pesticide Calibration Mix B (cat.# 32278) | |
Diluent: | Hexane |
Conc.: | Total OPP custom standard mix (0.2µg/mL), 1µL US EPA Method 8141A custom standard mixes (200ng/mL) |
Injection | splitless (hold 1.0 min) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 180 °C at 10 °C/min (hold 2 min) to 300 °C at 18 °C/min (hold 3 min) |
Dead Time: | 0.78 min @ 100 °C |
Detector | FPD @ 280 °C |