Explosives by EPA Method 8095 on Rtx-TNT

Peaks
1.Nitrobenzene
2.2-Nitrotoluene
3.3-Nitrotoluene
4.4-Nitrotoluene
5.Nitroglycerine
6.1,3-Dinitrobenzene
7.2,6-Dinitrotoluene
8.2,4-Dinitrotoluene
Peaks
9.3,4-Dinitrotoluene (IS)
10.1,3,5-Trinitrobenzene
11.Trinitrotoluene
12.PETN
13.RDX
14.4-Amino-2,6-dinitrotoluene
15.3,5-Dinitroaniline
16.2-Amino-4,6-dinitrotoluene
17.Tetryl
image
GC_EV00406
ColumnRtx-TNT, 6 m, 0.53 mm ID, 1.50 µm (cat.# 12998)
Sample8095 CAlibration Mix A (cat.# 31607)
8095 CAlibration Mix B (cat.# 31608)
3,4-dinitrotoluene (cat.# 31452)
Conc.:200-1,000 pg per compound, estimated
InjectionDirect
Liner:Siltek 1mm Uniliner (cat.# 21052-214.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:80 °C (hold 1 min) to 180 °C at 10 °C/min to 300 °C at 30 °C/min (hold 3 min)
Carrier GasHe
Dead Time:0.073 min
DetectorECD @ 330 °C
NotesHead pressure: 3 psi (20.7 kPa)
Flow rate: 17 mL/min. @ 80°C
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

EPA 8095

,

explosives

,

Rtx®-TNT

,

nitroaromatic

,

TNT

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.