Chlorophenoxyacid Herbicides US EPA Method 615 on Rtx-35
Peaks | |
---|---|
1. | Dalapon![]() |
2. | DCAA (S)![]() |
3. | Dicamba![]() |
4. | MCPP![]() |
5. | MCPA![]() |
6. | DBOB (IS)![]() |
7. | Dichlorprop![]() |
8. | 2,4-D![]() |
9. | 2,4,5-TP![]() |
10. | 2,4,5-T![]() |
11. | Dinoseb![]() |
12. | 2,4-DB![]() |

Column | Rtx-35, 30 m, 0.53 mm ID, 0.50 µm (cat.# 10440) |
Sample | Chlorophenoxy herbicides |
Conc.: | on-column concentration 10–10,000µg/mL |
Injection | |
Inj. Vol.: | 0.5 µL Direct |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 60 °C to 150 °C at 8 °C/min (hold 5 min) to 210 °C at 4 °C/min |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 35 cm/sec @ 60 °C |
Detector | ECD @ 275 °C |
Notes | ECD with anode purge |