Chlorophenoxyacid Herbicides US EPA Method 8151 Rtx-CLPesticides & Rtx-CLPesticides2

Peaks
1.Dalapon
2.3,5-Dichlorobenzoic acid
3.4-Nitroanisole
4.Dicamba
5.MCPP
6.MCPA
7.Dichlorprop
8.2,4-D
9.Pentachloroanisole
Peaks
10.2,4,5-TP
11.Chloramben
12.2,4,5-T
13.2,4-DB
14.Picloram
15.Bentazon
16.Dinoseb
17.DCPA methyl ester (Chlorthal-dimethyl)
18.Acifluorfen
image
GC_EV00091
ColumnsRtx-CLPesticides 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139)
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
SampleDerivatized chlorophenoxy herbicides
Conc.: On-column conc.: 200–20,000 pg/mL
Injection
Inj. Vol.:1.0 µL Direct
Liner:Uniliner® inlet liner (cat.# 20335)
Inj. Temp.:200 °C
Oven
Oven Temp.:80 °C (hold 1 min) to 300 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant pressure (12.5 psi, 86.2 kPa)
Temp.:80 °C
DetectorECD w/anode purge @ 300 °C
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Pesticides

,

Dual Column

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us