Nitrogen-Containing Herbicides on Rtx-5
Peaks | |
---|---|
1. | Eptam®![]() |
2. | Sutan®![]() |
3. | Vernam®![]() |
4. | Tillam®![]() |
5. | Ordram®![]() |
6. | Propachlor![]() |
7. | Ro-Neet®![]() |
8. | Treflan®![]() |
9. | Balan®![]() |
10. | Simazine![]() |
11. | Atrazine![]() |
Peaks | |
---|---|
12. | Propazine![]() |
13. | Tolban®![]() |
14. | Terbacil![]() |
15. | Sencor®![]() |
16. | Bromacil![]() |
17. | Dual®![]() |
18. | Paarlan®![]() |
19. | Prowl®![]() |
20. | Oxadiazon![]() |
21. | GOAL®![]() |
22. | Hexazinone![]() |

Column | Rtx-5, 30 m, 0.53 mm ID, 0.50 µm (cat.# 10240) |
Sample | Nitrogen-Containing Herbicides |
Conc.: | 10 ng per component (approx.) |
Injection | |
Inj. Vol.: | 0.2 µL Direct |
Liner: | Uniliner inlet liner |
Inj. Temp.: | 290 °C |
Oven | |
Oven Temp.: | 60 °C (hold 1 min) to 290 °C at 15 °C/min (hold 5 min) |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 40 cm/sec |
Detector | FID @ 290 °C |
Notes | Flow rate: 5.2 cc/min FID sensitivity: 16 x 10-11 AFS |