Phenols by EPA Method 604 on Rtx-200
Peaks | |
---|---|
1. | Phenol![]() |
2. | 2-Chlorophenol![]() |
3. | 2,4-Dimethylphenol![]() |
4. | 2,4-Dichlorophenol![]() |
5. | 2-Nitrophenol![]() |
Peaks | |
---|---|
6. | 4-Chloro-3-methylphenol![]() |
7. | 2,4,6-Trichlorophenol![]() |
8. | 4-Nitrophenol![]() |
9. | Pentachlorophenol![]() |
10. | 2,4-Dinitrophenol![]() |
11. | 2-Methyl-4,6-dinitrophenol![]() |

GC_EV00028
Column | Rtx-200, 30 m, 0.32 mm ID, 0.25 µm (cat.# 15024) |
Sample | 604 Phenols Calibration Mix (cat.# 31029) |
Conc.: | 200 ng |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 40:1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 50 °C (hold 4 min) to 250 °C at 6 °C/min |
Carrier Gas | He, constant pressure |
Linear Velocity: | 20 cm/sec |
Detector | FID @ 250 °C |
Notes | FID sensitivity: 4 x 10-11 AFS |