Formaldehyde on Rt-U-BOND

Peaks
1.Air
2.Carbon dioxide
3.Formaldehyde
4.Water
5.Methanol
image
GC_CH01137
ColumnRt-U-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19750)
SampleFormaldehyde (manual headspace)
Injection
Inj. Vol.:10 µL split (split ratio 10:1)
Liner:2 mm split Precision liner w/wool (cat.# 20823)
Inj. Temp.:200 °C
Split Vent Flow Rate:40 mL/min
Oven
Oven Temp.:100 °C (hold 1 min) to 150 °C at 25 °C/min (hold 3 min)
Carrier GasHe, constant pressure (7.7 psi, 53.1 kPa)
Linear Velocity:39 cm/sec @ 100 °C
DetectorTCD @ 200 °C
Make-up Gas Type:He
Data Rate:20 Hz
Sensitivity Mode:He/H2
InstrumentHP5890 GC
NotesSample Preparation: 10μl of formaldehyde solution added to a 20ml vial and heated to 60°C with CO2 added to the headspace. The formaldehyde solution is at a nominal concentration of 37% by weight with 10 to 15% methanol and the remaining amount water to stabilize the formaldehyde.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Formaldehyde

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us