Solvents on Rtx-5

Peaks
1.Acetone
2.Acrylonitrile
3.carbon disulfide impurity
4.Hexane
5.Methyl ethyl ketone
6.Ethyl acetate
7.Methyl Cellosolve®
8.Tetrahydrofuran
9.n-Butyl alcohol
Peaks
10.Isopropyl acetate
11.Heptane
12.n-Propyl acetate
13.Amyl alcohol
14.Toluene
15.n-Butyl acetate
16.n-Butyl acrylate
17.Cellosolve® acetate
18.Cyclohexanone
image
GC_CH00346
ColumnRtx-5, 60 m, 0.32 mm ID, 3.00 µm (cat.# 10287)
Samplesolvents collected on a charcoal tube and extracted in carbon disulfide
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:300 °C
Oven
Oven Temp.:50 °C (hold 4 min) to 120 °C at 5 °C/min (hold 20 min)
Carrier GasH2, constant pressure
Linear Velocity:40 cm/sec @ 50 °C
DetectorFID @ 300 °C
NotesFID sensitivity: 128 x 10-12 AFS
AcknowledgementPermission to publish this chromatogram granted by Kenda McCoy, Rexham Industrial.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Solvents

,

Rtx-5

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us