Solvents on Rtx-1
Peaks | |
---|---|
1. | Methanol![]() |
2. | Methyl formate![]() |
3. | Acetone![]() |
4. | Pentane![]() |
5. | Methylene chloride![]() |
6. | n-Propyl alcohol![]() |
7. | Methyl ethyl ketone![]() |
8. | sec-Butyl alcohol![]() |
9. | Ethyl acetate![]() |
10. | Chloroform![]() |
Peaks | |
---|---|
11. | Hexane![]() |
12. | Tetrahydrofuran![]() |
13. | 1,1,1-Trichloroethane![]() |
14. | Isobutyl alcohol![]() |
15. | Benzene![]() |
16. | n-Butyl alcohol![]() |
17. | Ethoxyethane![]() |
18. | Toluene![]() |
19. | n-Butyl acetate![]() |
20. | o-Xylene![]() |

Column | Rtx-1, 30 m, 0.53 mm ID, 5.00 µm (cat.# 10179) |
Sample | Solvent mix |
Injection | Direct |
Liner: | Uniliner liner |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C (hold 4 min) to 110 °C at 4 °C/min to 175 °C at 15 °C/min (hold 5 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 80 cm/sec |
Detector | FID @ 250 °C |
Notes | 0.5 mg/mL wet needle injection Flow rate: 10 cc/min FID sensitivity: 64 x 10-11 AFS |