Solvent Mixture #2 on Rtx-1

Peaks
1.Pentane
2.Methylene Chloride
3.Ethylene glycol
4.Heptane
5.Cyclopentanol
6.3-Hexanol
7.Acetamide
8.2-Methyl-1-pentanol
9.Furfuryl alcohol
10.Butyl ether
11.Nonane
12.Cumene
13.Ethyl amyl ketone
14.Heptanol
15.Butyl butanoate
16.Unknown
17.Benzyl Alcohol
18.Dipropylene glycol
19.benzene, diethyl-
20.Unknown
21.Unknown
22.Hexachloroethane
23.Undecane
24.1-Nonanol
25.p-Methoxyphenol
26.Triethylene glycol
27.Dodecane
28.Undecanal
29.Tridecane
30.Unknown
31.Dodecanal
32.Dicyclohexylamine
33.bis(2,2-methoxy)ethyl ether
34.Pentadecane
35.Heptadecane
36.Octadecane
37.Nonadecane
38.Eicosane
39.Acetyl tributyl citrate
40.2-Buten-1-ol
41.Formamide
42.3-Pentanol
43.1-Nitropropane
44.Dimethylformamide
45.2-Methyl-3-pentanol
46.Toluene
47.Ethyl chloroacetate
48.Dimethylacetamide
49.p-Xylene
50.sec-tetrachloroethane
51.Benzaldehyde
Peaks
52.o-Chlorotoluene
53.2,6-Dimethyl-4-heptanone
54.2-Octanone
55.o-Cresol
56.α-Methylbenzyl alcohol
57.5-Nonanone
58.Nonanal
59.Decanal
60.Unknown
61.1-Decanol
62.1-Undecanol
63.2-Dodecanone
64.1-Dodecanol
65.Tetraethylene glycol
66.Dibenzyl
67.Diethyl Phthalate
68.Tributyl phosphate
69.Diphenyl sulfone
70.Allyl alcohol
71.Unknown
72.Isopropyl acetate
73.Benzene
74.2-Nitropropane
75.Nitroethane
76.Pentanal
77.2-Bromobutane
78.1-Chloropentane
79.Cyclopentanone
80.2-Hexanol
81.Butyl acetate
82.2-Ethyl-1-butanol
83.3-Ethyl-3-pentanol
84.1,4-Dichlorobutane
85.2-Methyl-2,4-pentanediol
86.Butoxyethanol
87.1,2,3-Trichloropropane
88.1,4-Butanediol
89.Methyl hexanoate
90.1,2,4-Trimethylbenzene
91.2-Ethyl-1-hexanol
92.Dipentene/Limonene
93.Tetrahydrofurfuryl acetate
94.Unknown
95.Decahydronaphthalene
96.Unknown
97.Unknown
98.2-Decanol
99.1,2-Bis(2-methoxyethoxy)ethane
100.2-Phenoxyethanol
101.Unknown
102.Benzyl ether
image
GC_CH00338
ColumnRtx-1, 60 m, 0.53 mm ID, 3.00 µm (cat.# 10188)
SampleSolvent mix #2
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:275 °C
Split Vent Flow Rate:50 mL/min
Oven
Oven Temp.:40 °C (hold 5 min) to 285 °C at 5 °C/min
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorMS
Mode:Scan
Source Temp.:285 °C