Questo cromatogramma potrebbe contenere analiti che sono sinonimi dei vostri parametri di ricerca.

Derivatized Cannabinoids (50 ng/mL) on Rxi-5Sil MS

image
GC_CF1146
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Ethyl acetate
Conc.:50 ng/mL TMS derivatives
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5 mm splitless taper w/wool (cat.# 22286-200.1)
Inj. Temp.:250 °C
Purge Flow:21.4 mL/min
Oven
Oven Temp.:150 °C to 330 °C at 15 °C/min (hold 3 min)
Carrier GasHe, constant linear velocity
Linear Velocity:40 cm/sec, 13.8 psi, 95.1 kPa @ 150 °C
DetectorMS
Mode:SIM
SIM Program:390, 337, 458, 367, 382, 310, 371, 386, 343, 473, 398 m/z
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Solvent Delay Time:4 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Cannabinoids

,

GCMS

,

Rxi-5

,

Rxi-5MS

,

Rxi-5SilMS

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.