Innovative Chromatography Products
www.restek.it
www.restek.it

Derivatized Cannabinoids (50 ng/mL) on Rxi-5Sil MS

GC_CF1146
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
Sample | |
Diluent: | Ethyl acetate |
Conc.: | 50 ng/mL TMS derivatives |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | 3.5 mm splitless taper w/wool (cat.# 22286-200.1) |
Inj. Temp.: | 250 °C |
Purge Flow: | 21.4 mL/min |
Oven | |
Oven Temp.: | 150 °C to 330 °C at 15 °C/min (hold 3 min) |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 40 cm/sec, 13.8 psi, 95.1 kPa @ 150 °C |
Detector | MS |
Mode: | SIM |
SIM Program: | 390, 337, 458, 367, 382, 310, 371, 386, 343, 473, 398 m/z |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 200 °C |
Solvent Delay Time: | 4 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Instrument | Shimadzu 2010 GC & QP2010+ MS |
relativo alle ricerche nella libreria dei cromatogrammi
Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.