Derivatized Amphetamines (500 ng/mL) on Rxi-5Sil MS

Peaks
1.Amphetamine
2.Methamphetamine
Peaks
3.MDA
4.MDMA
5.MDEA
c = contaminant
image
GC_CF1144
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Butyl chloride
Conc.:500 ng/mL HFAA derivatives
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5 mm splitless taper w/wool (cat.# 22286-200.1)
Inj. Temp.:250 °C
Purge Flow:28 mL/min
Oven
Oven Temp.:75 °C to 300 °C at 15 °C/min
Carrier GasHe, constant linear velocity
Linear Velocity:45 cm/sec, 13.5 psi, 93.1 kPa @ 75 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:4 min
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:40-300 amu
Scan Rate:5 scans/sec
InstrumentShimadzu 2010 GC & QP2010+ MS

relativo alle ricerche nella libreria dei cromatogrammi

Amphetamines

,

GCMS

,

Rxi-5

,

Rxi-5MS

,

Rxi-5SilMS

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.


Restek Domestic Customer Service

Oggetto

Messaggio

Nome e Cognome

Vostro indirizzo e-mail

Azienda

Indirizzo

Spam Block (Please leave this blank)
Aggiungi allegato, anche in Drag&Drop (facoltativo)
...

Campi Obbligatori

Grazie.

Il messaggio è stato inviato. La ricontatteremo non appena possibile.

Messaggio non inviato.

Siamo spiacenti, al momento non è stato possibile inviare il suo messaggio. Preghiamo di riprovare più tardi oppure di contattarci telefonicamente al numero +39 027610037

www.restek.com/Contact-Us