Comparison of Converted and Unconverted GHB on Rtx-BAC1
Peaks | |
---|---|
1. | GBL![]() |
Peaks | |
---|---|
2. | AMGB (IS)![]() |

GC_CF01039
Column | Rtx-BAC1, 30 m, 0.32 mm ID, 1.80 µm (cat.# 18003) |
Sample | |
Conc.: | 50 µg/mL GHB, GBL, and AMGB (IS) in urine A: unconverted B: converted with sulfuric acid |
Injection | |
Inj. Vol.: | 1000 µL headspace-syringe split (split ratio 10:1) |
Liner: | 1mm Split (cat.# 20972) |
Inj. Temp.: | 200 °C |
Headspace-Syringe | |
Sample Temp.: | 100 °C |
Sample Equil. Time: | 10 min |
Oven | |
Oven Temp.: | 50 °C (hold 3 min) to 150 °C at 20 °C/min (hold 7 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.0 mL/min |
Detector | FID @ 240 °C |
Make-up Gas Flow Rate: | 40 mL/min |