Cannabinoids Rxi-5ms

Peaks
1.Cannabidiol
2.Δ8-Tetrahydrocannabinol
3.Δ9-Tetrahydrocannabinol
Peaks
4.Cannabinol
5.11-Hydroxy-Δ9-tetrahydrocannabinol
6.11-Nor-Δ9-tetrahydrocannabinol carboxylic acid
image
GC_CF00891
ColumnRxi-5ms, 12 m, 0.20 mm ID, 0.33 µm (cat.# 13497)
Sample
Diluent:Methanol
Conc.:1000 µg/mL each component
Injection
Inj. Vol.:1.0 µL split (split ratio 25:1)
Liner:Single taper (4mm), base deactivated w/base deactivated wool
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C to 340 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Ionization Mode:EI
Scan Range:100-550 amu