Agilent GCs
Shimadzu GCs
Thermo GCs
Varian GCs
PerkinElmer GCs
SRI and Buck GCs


Agilent GC Nut and Ferrule Compatibility

Available Ferrule Cat.#

Description

Compatible Nut Cat.#

22214, 20200, 20227, 20211, 20229, 27062, 27063

Standard Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

20883 & 21879



Capillary Column Nuts

21041


Finger-Tight Capillary Column Nuts

21312


Finger-Tight Capillary Column Nuts

22633 & 22636



Extended Capillary Column Nuts

22347



Hot Swap Capillary Column Nuts

22215, 20201, 20228, 20212, 20231, 27064, 27065

Standard Ferrules
(Fit 0.32 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.25/0.32 mm ID
MXT Columns)

22216, 20202, 20224, 20213, 20230, 27067, 27068

Standard Ferrules
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.53 mm ID
MXT Columns)

20232, 27066

Standard Ferrules
(Fit 0.28 mm ID
MXT Columns)

Capillary Ferrules (includes all the above)

Capillary Ferrules
for 1/16" Compression-Type Fittings

20250, 20251, 20238, 20239, 27070, 27071

Compact Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

21884 & 21878




Capillary Column Nuts

21040



Finger-Tight Capillary Column Nuts

21311



Finger-Tight Capillary Column Nuts

22634 & 22637



Extended Capillary Column Nuts

22348



Hot Swap Capillary Column Nuts

21007, 21008, 20248, 20249, 27072, 27073

Compact Ferrules
(Fit 0.32 mm ID Fused Silica Columns)

&
Compact Ferrules
(Fit 0.25/0.32 mm ID
MXT Columns)

20252, 20253, 20263, 20264, 27074, 27075

Compact Ferrules
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Compact Ferrules
(Fit 0.53 mm ID
MXT Columns)

21059, 21061

Compact Ferrules
for Micropacked Columns

Need more help finding the correct ferrule to install your column in an Agilent GC?

Available Ferrule Cat.#

Description

Compatible Nut Cat.#

22214, 20200, 20227, 20211, 20229, 27062, 27063

Standard Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

20883 & 21879
Capillary Column Nuts

21041
Finger-Tight Capillary Column Nuts

21312
Finger-Tight Capillary Column Nuts

22633 & 22636
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

22215, 20201, 20228, 20212, 20231, 27064, 27065

Standard Ferrules
(Fit 0.32 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.25/0.32 mm ID
MXT Columns)

20883 & 21879
Capillary Column Nuts

21041
Finger-Tight Capillary Column Nuts

21312
Finger-Tight Capillary Column Nuts

22633 & 22636
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

22216, 20202, 20224, 20213, 20230, 27067, 27068

Standard Ferrules
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.53 mm ID
MXT Columns)

20883 & 21879
Capillary Column Nuts

21041
Finger-Tight Capillary Column Nuts

21312
Finger-Tight Capillary Column Nuts

22633 & 22636
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

20232, 27066

Standard Ferrules
(Fit 0.28 mm ID
MXT Columns)

20883 & 21879
Capillary Column Nuts

21041
Finger-Tight Capillary Column Nuts

21312
Finger-Tight Capillary Column Nuts

22633 & 22636
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

Capillary Ferrules (includes all the above)

Capillary Ferrules
for 1/16" Compression-Type Fittings

20883 & 21879
Capillary Column Nuts

21041
Finger-Tight Capillary Column Nuts

21312
Finger-Tight Capillary Column Nuts

22633 & 22636
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

20250, 20251, 20238, 20239, 27070, 27071

Compact Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

21884 & 21878
Capillary Column Nuts

21040
Finger-Tight Capillary Column Nuts

21311
Finger-Tight Capillary Column Nuts

22634 & 22637
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

21007, 21008, 20248, 20249, 27072, 27073

Compact Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

21884 & 21878
Capillary Column Nuts

21040
Finger-Tight Capillary Column Nuts

21311
Finger-Tight Capillary Column Nuts

22634 & 22637
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

20252, 20253, 20263, 20264, 27074, 27075

Compact Ferrules
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Compact Ferrules
(Fit 0.53 mm ID
MXT Columns)

21884 & 21878
Capillary Column Nuts

21040
Finger-Tight Capillary Column Nuts

21311
Finger-Tight Capillary Column Nuts

22634 & 22637
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

21059, 21061

Compact Ferrules
for Micropacked Columns

21884 & 21878
Capillary Column Nuts

21040
Finger-Tight Capillary Column Nuts

21311
Finger-Tight Capillary Column Nuts

22634 & 22637
Extended Capillary Column Nuts

22348
Hot Swap Capillary Column Nuts

Need more help finding the correct ferrule to install your column in an Agilent GC?

Back to top of page


Shimadzu GC Nut and Ferrule Compatibility

Back to top of page


Thermo GC Nut and Ferrule Compatibility

Available Ferrule Cat.#

Description

Compatible Nut Cat.#

22214, 20200, 20227, 20211, 20229, 27062, 27063

Standard Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

24916

Adaptor For Capillary Column on Detector Base

24917


Adaptor for Capillary Column
on Split/Splitless Injector

20543

Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

23150

Stainless Steel 1/16" Nut

22215, 20201, 20228, 20212, 20231, 27064, 27065

Standard Ferrules
(Fit 0.32 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.25/0.32 mm ID MXT Columns)

22216, 20202, 20224, 20213, 20230, 27067, 27068

Standard Ferrules
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.53 mm ID
MXT Columns)

20232, 27066

Standard Ferrules
(Fit 0.28 mm ID
MXT Columns)

Capillary Ferrules
(includes all the above)

Capillary Ferrules for 1/16" Compression-Type Fittings

20280, 20281

Ferrules for Thermo Scientific GCs
(Fit ≤ 0.28 mm ID Fused Silica Columns)

24969

Adaptor for Capillary Column on Detector Base

24970

Adaptor for Capillary Column on Split/Splitless Injector

20544

Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

24973

Fixing Nut for Capillary Column for Split/Splitless Injector on Thermo Scientific TRACE, 8000, 8000 TOP & Focus SSL

20282, 20283

Ferrules for Thermo Scientific GCs
(Fit 0.32 mm ID Fused Silica Columns)

&
Ferrules for Thermo Scientific GCs (Fit 0.25/0.32 mm ID
MXT Columns)

20284, 20285

Ferrules for Thermo Scientific GCs
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Ferrules for Thermo Scientific GCs (Fit 0.45/0.53 mm ID
MXT Columns)

Need more help finding the correct ferrule to install your column in a Thermo Scientific GC?

Available Ferrule Cat.#

Description

Compatible Nut Cat.#

22214, 20200, 20227, 20211, 20229, 27062, 27063

Standard Ferrules
(Fit ≤ 0.28 mm ID Fused Silica Columns)

24916
Adaptor For Capillary Column on Detector Base

24917
Adaptor for Capillary Column on Split/Splitless Injector

20543
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

23150
Stainless Steel 1/16" Nut

22215, 20201, 20228, 20212, 20231, 27064, 27065

Standard Ferrules
(Fit 0.32 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.25/0.32 mm ID MXT Columns)

24916
Adaptor For Capillary Column on Detector Base

24917
Adaptor for Capillary Column on Split/Splitless Injector

20543
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

23150
Stainless Steel 1/16" Nut

22216, 20202, 20224, 20213, 20230, 27067, 27068

Standard Ferrules
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Standard Ferrules
(Fit 0.53 mm ID
MXT Columns)

24916
Adaptor For Capillary Column on Detector Base

24917
Adaptor for Capillary Column on Split/Splitless Injector

20543
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

23150
Stainless Steel 1/16" Nut

20232, 27066

Standard Ferrules
(Fit 0.28 mm ID
MXT Columns)

24916
Adaptor For Capillary Column on Detector Base

24917
Adaptor for Capillary Column on Split/Splitless Injector

20543
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

23150
Stainless Steel 1/16" Nut

Capillary Ferrules
(includes all the above)

Capillary Ferrules for 1/16" Compression-Type Fittings

24916
Adaptor For Capillary Column on Detector Base

24917
Adaptor for Capillary Column on Split/Splitless Injector

20543
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

23150
Stainless Steel 1/16" Nut

20280, 20281

Ferrules for Thermo Scientific GCs
(Fit ≤ 0.28 mm ID Fused Silica Columns)

24969
Adaptor for Capillary Column on Detector Base

24970
Adaptor for Capillary Column on Split/Splitless Injector

20544
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

24973
Fixing Nut for Capillary Column for Split/Splitless Injector on Thermo Scientific TRACE, 8000, 8000 TOP & Focus SSL

20282, 20283

Ferrules for Thermo Scientific GCs
(Fit 0.32 mm ID Fused Silica Columns)

&
Ferrules for Thermo Scientific GCs (Fit 0.25/0.32 mm ID
MXT Columns)

24969
Adaptor for Capillary Column on Detector Base

24970
Adaptor for Capillary Column on Split/Splitless Injector

20544
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

24973
Fixing Nut for Capillary Column for Split/Splitless Injector on Thermo Scientific TRACE, 8000, 8000 TOP & Focus SSL

20284, 20285

Ferrules for Thermo Scientific GCs
(Fit 0.45/0.53 mm ID Fused Silica Columns)

&
Ferrules for Thermo Scientific GCs (Fit 0.45/0.53 mm ID
MXT Columns)

24969
Adaptor for Capillary Column on Detector Base

24970
Adaptor for Capillary Column on Split/Splitless Injector

20544
Siltek-Treated Adaptor for Capillary Column on Split/Splitless Injector

24973
Fixing Nut for Capillary Column for Split/Splitless Injector on Thermo Scientific TRACE, 8000, 8000 TOP & Focus SSL

Need more help finding the correct ferrule to install your column in a Thermo Scientific GC?

Back to top of page


Varian GC Nut and Ferrule Compatibility

Back to top of page


PerkinElmer GC Nut and Ferrule Compatibility

Back to top of page


SRI and Buck GC Nut and Ferrule Compatibility

Back to top of page