GCxGC Analysis of Complex Petroleum Hydrocarbons: Sulfur Speciation in Diesel

By Jack Cochran and Jan Pijpelink

Content previously published in Petro Industry News

Comprehensive 2D GC, also known as GCxGC, is a powerful technique with a great deal of potential for improving separations of complex petroleum and petrochemical samples that contain hundreds—or even thousands—of components. While GCxGC is still considered emerging technology and is used primarily in research and development labs, it has undergone significant growth in the past few years.

As evidenced by recent literature, industrial applications for GCxGC are growing in a number of areas; however, its ability to increase peak capacity and advance separations in complex matrices makes it particularly useful in the petro industries. GCxGC can be used to analyze hydrocarbons ranging in volatility from C4 (butane) to C40 (tetracontane), although with automatic flow programming of the cold jet in the thermal modulator analyses up to C48 have been reported [1]. Other notable recent advances include a full characterization of middle distillates using supercritical fluid chromatography [2], a high temperature analysis of vacuum gas oils up to nC60 [3], and a partial characterization of military fog oil [4]. Military fog oil is an extremely complex middle distillate product containing mainly aliphatic compounds. Using GCxGC-TOFMS, Kohl et al. were able to identify a wide range of low concentration aromatics from what would typically be an unresolved complex mixture. Compounds reported included alkanes, cyclohexanes, hexahydroindanes, decalins, adamantanes, bicyclohexanes, alkylbenzenes, indanes, tetrahydronaphthalenes, partially hydrogenated polycyclic aromatic hydrocarbons, biphenyls, dibenzofurans, and dibenzothiophenes. In these and other examples, the analytical power of comprehensive 2D GC is used to achieve a more advanced separation of complex matrices than is possible with 1D GC.

Technical Overview of GCxGC

In comprehensive 2D GC, two independent separations are applied to a single sample injection. A summary of the technique is given here; however, recent review papers provide a more complete description, focusing on recent advances and industrial applications [5,6]. In GCxGC, the first separation is usually based on boiling point and uses a standard nonpolar phase. Next, a thermal or valve modulator is used to focus the effluent from the first column onto the second column, which is a short (1-2 m) column and typically is a polar phase. Inverse polarity column set-ups are also sometimes employed, but with either approach the key to maximizing use of the separation space is to choose orthogonal columns that differ significantly in selectivity. Separation results are displayed as a contour plot, which is a three dimensional representation of intensity (z) across the retention times of both column 1 (x) and column 2 (y). The result is a technique with increased peak capacity that can be used to separate compounds which coelute in the first dimension. Ultimately, the increased peak capacity obtainable with GCxGC allows better characterization of complex petro fractions such as naphtha, gasoline, and diesel.

Sulfur Speciation in Diesel

The analysis of sulfurs, such as dibenzothiophenes, in diesel offers a timely example of the benefits of GCxGC for complex hydrocarbon samples. In recent years, new vehicle emissions standards have been adopted across the globe; these requirements have driven the development of ultra low sulfur diesel that is compatible with new emission control technology. In many countries, specifications for total sulfur in diesel have been reduced from 50 ppm to 10-15 ppm, making the analysis of various sulfur components even more crucial for the refinery industry. Using the analysis of dibenzothiophenes as an example, it is clear that 1D GC is insufficient to distinguish sulfur compounds due to coelutions with other components in the diesel matrix (Figure 1). An MS detector was used for this work because it offers the potential to identify a broad range of analytes from a single injection, compared to the limited information that can be obtained using a sulfur-specific detector. However, even with the power of the MS, the diesel sample was too complex to yield usable results. The spectrum taken at the retention time for dibenzothiophene confirms that the separation of components is not sufficient and the sulfur compounds cannot be identified (Figure 2).

Figure 1:  Elution area of dibenzothiophenes in a one-dimensional GC analysis of diesel. The sample is so complex that no distinct chromatographic peaks can be seen.

One-dimensional GC-TOFMS chromatogram of the dibenzothiophene elution area for a diesel sample.
GC_PC1208
ColumnRxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
  with Rxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
SampleDiesel fuel #2 composite (cat.# 31093)
Diluent:Methylene chloride
Conc.:5,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:100 mL/min
Oven
Oven Temp.:Rxi-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min)
:
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:0 sec
Hot Pulse Time:0 sec
Cool Time between Stages:0 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:50 mu/100 u
Solvent Delay Time:10 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 450 amu
Spectral Acquisition Rate:2 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesTwo 60 m columns were connected using a Press-Tight connector to make a 120 m analytical column.

Figure 2:  Mass spectrum taken at dibenzothiophene retention time for one-dimensional GC of diesel. The mass spectrum is not representative of dibenzothiophene at all due to sample complexity.

Mass spectrum taken at dibenzothiophene retention time for one-dimensional GC-TOFMS.
GC_PC1207
ColumnRxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
  with Rxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
SampleDiesel fuel #2 composite (cat.# 31093)
Diluent:Methylene chloride
Conc.:5,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:100 mL/min
Oven
Oven Temp.:Rxi-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min)
:
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:0 sec
Hot Pulse Time:0 sec
Cool Time between Stages:0 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:50 mu/100 u
Solvent Delay Time:10 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 450 amu
Spectral Acquisition Rate:2 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS

In contrast, when GCxGC is used for the analysis of sulfurs in diesel, individual dibenzothiophenes are resolved both from the interferences that obscured them in the one-dimensional analysis and also from each other. The structured chromatogram (contour plot) that is obtained offers good separation of major groups and even resolves multiple positional isomers within each carbon substituent group (Figures 3-5). Comparison of the sample mass spectrum at the retention time for dibenzothiophene to a reference spectrum reveals a clear match and confirms compound identification (Figure 6). As shown in this example, good use of the separation space is made by optimizing parameters for both columns. Often, in order to optimize separations on the second column, the first dimension separation is intentionally slowed in order to produce broader peaks that will accommodate 3-5 modulations per peak. With careful attention to flows, temperatures, and modulator settings, good column efficiencies can be obtained in both dimensions. By employing GCxGC and optimizing use of the separation space, peak capacity is significantly increased. This approach, in combination with the use of a powerful MS detector, allows speciation of sulfur compounds in diesel and provides a level of sample characterization that cannot be obtained using one-dimensional GC.

Analysis of sulfurs in diesel provides a good demonstration of the power of GCxGC for complex matrices. Increased peak capacity allows identification of relevant constituents that cannot be differentiated by GC alone. As this technology is increasingly adopted by the petroleum and petrochemical industries, better materials characterization and process optimization can be achieved.

Figure 3:  GCxGC-TOFMS chromatogram (contour plot) of diesel analyzed with orthogonal Rxi-1ms and Rtx-FLD columns showing structured separation of aliphatics and aromatics.

Diesel on Rxi-1ms and Rtx-FLD (GCxGC-TOFMS, TIC)
GC_PC1203
ColumnRxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
  with Rxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
Rtx-FLD 1.5 m, 0.25 mm ID, 0.10 µm,
SampleDiesel fuel #2 composite (cat.# 31093)
Diluent:Methylene chloride
Conc.:5,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:100 mL/min
Oven
Oven Temp.:Rxi-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min)
Rtx-FLD: 45 °C (hold 1 min) to 325 °C at 1.3 °C/min (hold 1 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:1 sec
Cool Time between Stages:0.5 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:50 mu/100 u
Solvent Delay Time:10 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 450 amu
Spectral Acquisition Rate:200 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesThe Rtx-FLD column is an experimental column and is not yet commercially available.

Figure 4:  GCxGC-TOFMS contour plot (TIC) of the dibenzothiophene elution region for a diesel sample. The dibenzothiophenes are more retained by the Rtx-FLD column, which means they are higher on the y-axis of the contour plot than most other components.

Dibenzothiophene (Including C1-C4 Substitution) Elution Area for a Diesel Sample on Rxi-1ms and Rtx-FLD (GCxGC-TOFMS, TIC)
GC_PC1204
ColumnRxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
  with Rxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
Rtx-FLD 1.5 m, 0.25 mm ID, 0.10 µm,
SampleDiesel fuel #2 composite (cat.# 31093)
Diluent:Methylene chloride
Conc.:5,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:100 mL/min
Oven
Oven Temp.:Rxi-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min)
Rtx-FLD: 45 °C (hold 1 min) to 325 °C at 1.3 °C/min (hold 1 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:1 sec
Cool Time between Stages:0.5 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:50 mu/100 u
Solvent Delay Time:10 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 450 amu
Spectral Acquisition Rate:200 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesThe Rtx-FLD column is an experimental column and is not yet commercially available.

Figure 5:  GCxGC-TOFMS extracted ion contour plot of the dibenzothiophene elution area for a diesel sample. Plotting m/z ions 184, 198, 212, 226, and 240 enhances visualization of the separation of the dibenzothiophenes (including the alkyl substituted isomers).

Dibenzothiophene (Including C1-C4 Substitution) Elution Area for a Diesel Sample on Rxi-1ms and Rtx-FLD (GCxGC-TOFMS, Extracted Ions)
GC_PC1205
ColumnRxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
  with Rxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
Rtx-FLD 1.5 m, 0.25 mm ID, 0.10 µm,
SampleDiesel fuel #2 composite (cat.# 31093)
Diluent:Methylene chloride
Conc.:5,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:100 mL/min
Oven
Oven Temp.:Rxi-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min)
Rtx-FLD: 45 °C (hold 1 min) to 325 °C at 1.3 °C/min (hold 1 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:1 sec
Cool Time between Stages:0.5 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:50 mu/100 u
Solvent Delay Time:10 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 450 amu
Spectral Acquisition Rate:200 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesThe Rtx-FLD column is an experimental column and is not yet commercially available.
Extracted Ions: 184, 198, 212, 226, 240

Figure 6:  Mass spectrum of dibenzothiophene for comprehensive two-dimensional GC of diesel. The sample mass spectrum (top) is an excellent match with the NIST library mass spectrum (bottom).

Mass spectrum of dibenzothiophene in a diesel sample after separation with GCxGC-TOFMS, compared to a NIST library mass spectrum of dibenzothiophene.
GC_PC1206
ColumnRxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
  with Rxi-1ms 60 m, 0.25 mm ID, 0.25 µm (cat.# 13326)
Rtx-FLD 1.5 m, 0.25 mm ID, 0.10 µm,
SampleDiesel fuel #2 composite (cat.# 31093)
Diluent:Methylene chloride
Conc.:5,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:100 mL/min
Oven
Oven Temp.:Rxi-1ms: 40 °C (hold 1 min) to 320 °C at 1.3 °C/min (hold 1 min)
Rtx-FLD: 45 °C (hold 1 min) to 325 °C at 1.3 °C/min (hold 1 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:1 sec
Cool Time between Stages:0.5 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:50 mu/100 u
Solvent Delay Time:10 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 450 amu
Spectral Acquisition Rate:200 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesThe Rtx-FLD column is an experimental column and is not yet commercially available.

References

[1] W. Rathbun, J. Chromatogr. Sci., 45 (2007) 636.
[2] F. Adam, D. Thiebaut, F. Bertoncini, M. Courtiade, M.-C. Hennion, J. Chromatogr. A, 1217 (2010) 1386.
[3] T. Dutriez, M. Courtiade, D. Thiebaut, H. Dulot, F. Bertoncinia, J. Vial, M.C. Hennion, J. Chromatogr. A, 1216 (2009) 2905.
[4] A. Kohl, J. Cochran, D.M. Cropek, J. Chromatogr. A, 1217 (2010) 550.
[5] F.L. Dorman, J.J. Whiting, J.W. Cochran, J. Gardea-Torresdey, Anal. Chem., 82 (2010) 4775.
[6] F.L. Dorman, E.B. Overton, J.J. Whiting, J.W. Cochran, J. Gardea-Torresdey, Anal. Chem., 80 (2008) 4487.